Boston Scientific Corporation CRM312005 Manuel D’Utilisation

Page de 50
23
No analog 
output
Using incorrect output 
port and/or 
connections
Recheck connection of the 
Model 6924 External Recorder 
Cable.
Patient data disk 
error
Using disk created for a 
previous model of PRM 
or unformatted floppy 
disk
Use only Guidant Model 6627 
Patient Data Disks.
Write-protect tab open
Close write-protect tab.
Noise problems: 
ECG
Improper patient 
connections
Recheck patient leads for 
adequate skin contact and 
correct limb lead placement.
Excessive radio 
emissions from 
equipment
Check surrounding area for 
electrical equipment that is 
powered on and not needed. 
Move unneeded equipment 
away from patient and/or PRM, 
or turn off unneeded 
equipment. Consult ECG 
textbooks for additonal ECG 
techniques.
Telemetry: no 
communication
Incorrect application 
software or incorrect 
PRM for pulse 
generator
Install proper application 
software for pulse generator in 
use.
Incomplete telemetry 
communication
Reposition wand over the 
device; repeat interrogation.
Telemetry: 
intermittent 
communication
Incorrect telemetry 
wand
Use only the Guidant 
Model 6577 Sterilizable 
Telemetry Wand.
Excessive radio 
emissions from 
equipment
Check surrounding area for 
electrical equipment that is 
powered on and not needed. 
Move unneeded equipment 
away from patient and/or PRM, 
or turn off unneeded 
equipment.
Incomplete telemetry 
communication
Reposition wand over the 
device; repeat interrogation.
Displayed clock 
does not 
consistently 
keep time after 
setting
Low battery
Return to Guidant for 
replacement of clock battery.
Table 2.  Possible Causes and Corrective Actions for PRM Problems
Symptom
Possible Cause
Corrective Action