Boston Scientific Corporation CRM312005 Manuel D’Utilisation
23
No analog
output
output
Using incorrect output
port and/or
connections
port and/or
connections
Recheck connection of the
Model 6924 External Recorder
Cable.
Model 6924 External Recorder
Cable.
Patient data disk
error
error
Using disk created for a
previous model of PRM
or unformatted floppy
disk
previous model of PRM
or unformatted floppy
disk
Use only Guidant Model 6627
Patient Data Disks.
Patient Data Disks.
Write-protect tab open
Close write-protect tab.
Noise problems:
ECG
ECG
Improper patient
connections
connections
Recheck patient leads for
adequate skin contact and
correct limb lead placement.
adequate skin contact and
correct limb lead placement.
Excessive radio
emissions from
equipment
emissions from
equipment
Check surrounding area for
electrical equipment that is
powered on and not needed.
Move unneeded equipment
away from patient and/or PRM,
or turn off unneeded
equipment. Consult ECG
textbooks for additonal ECG
techniques.
electrical equipment that is
powered on and not needed.
Move unneeded equipment
away from patient and/or PRM,
or turn off unneeded
equipment. Consult ECG
textbooks for additonal ECG
techniques.
Telemetry: no
communication
communication
Incorrect application
software or incorrect
PRM for pulse
generator
software or incorrect
PRM for pulse
generator
Install proper application
software for pulse generator in
use.
software for pulse generator in
use.
Incomplete telemetry
communication
communication
Reposition wand over the
device; repeat interrogation.
device; repeat interrogation.
Telemetry:
intermittent
communication
intermittent
communication
Incorrect telemetry
wand
wand
Use only the Guidant
Model 6577 Sterilizable
Telemetry Wand.
Model 6577 Sterilizable
Telemetry Wand.
Excessive radio
emissions from
equipment
emissions from
equipment
Check surrounding area for
electrical equipment that is
powered on and not needed.
Move unneeded equipment
away from patient and/or PRM,
or turn off unneeded
equipment.
electrical equipment that is
powered on and not needed.
Move unneeded equipment
away from patient and/or PRM,
or turn off unneeded
equipment.
Incomplete telemetry
communication
communication
Reposition wand over the
device; repeat interrogation.
device; repeat interrogation.
Displayed clock
does not
consistently
keep time after
setting
does not
consistently
keep time after
setting
Low battery
Return to Guidant for
replacement of clock battery.
replacement of clock battery.
Table 2. Possible Causes and Corrective Actions for PRM Problems
Symptom
Possible Cause
Corrective Action