Boston Scientific Corporation CRM312005 Manuel D’Utilisation

Page de 50
39
Table 7.  Recommended Separation Distances between Portable and Mobile 
RF Communications Equipment and the Model 3120 PRM
a,
 
b,
 
c,
 
d
a.  The Model 3120 PRM is intended for use in an electromagnetic environment in which 
radiated RF disturbances are controlled. The customer or the user of the Model 3120 PRM 
can help prevent electromagnetic interference by maintaining a minimum distance between 
portable and mobile RF communications equipment (transmitters) and the Model 3120 PRM 
as recommended above, according to the maximum output power of the communications 
equipment.
b.  For transmitters rated at a maximum output power not listed above, the recommended 
separation distance d in meters (m) can be estimated using the equation applicable to the 
frequency of the transmitter, where P is the maximum output power rating of the transmitter 
in watts (W) according to the transmitter manufacturer.
c.  At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
d.  These guidelines may not apply in all situations. Electromagnetic propagation is affected by 
absorption and reflection from structures, objects, and people.
Rated maximum 
output power of 
transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 
80 MHz
80 MHz to 
800 MHz
800 MHz to 
2.5 GHz
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.70
3.70
7.38
100
11.70
11.70
23.33
d
1.17 P
=
d
1.17 P
=
d
2.33 P
=