Boston Scientific Corporation CRM312005 Manuel D’Utilisation

Page de 50
40
SPECIFICATIONS (Nominal)
Table 8. Nominal Specifications
Safety classification
Class I, Type CF, Defibrillation protected
Dimensions
18.5 in (47 cm) deep, 14.5 in (36.8 cm) wide, 
4.75 in (12.1 cm) high
Weight (approximate)
21.5 lb (9.8 kg)
Power rating
100–240 V, 50/60 Hz, 3.8/1.9 A
Power cord
8 ft (2.4 m), 100–240 V. Reliable grounding is 
achieved only when equipment is connected 
to receptacle marked with “Hospital only” or 
“Hospital grade.”
Operating temperature
50
°F to 105°F (10°C to 35°C)
Transport and storage 
temperature
–40
°F to 160°F (–40°C to 70°C)
Operating, transport, and 
storage humidity
25% to 95%
Atmospheric pressure
7.252 psi to 15.374 psi (50 kPa to 106 kPa)
External printer support
DB 25 parallel port connector
External monitor support
DB 15 VGA port connector
Analog output
+ 1 V output via seven-pin DIN connector
ECG performance
• Minimum amplitude 
detected
4.56
µV
• Lead selection
I, II, III, aVR, aVL, aVF, V
• Intrinsic ventricular rate 
display
30 bpm to 240 bpm + 6 bpm
• Input impedance
> 2.5 M 
Ω
• Electrode offset tolerance
300 mV
• Storage resolution
800 samples/sec, 4.56 
µV
• Filter settings
ON: 0.5 Hz to 25 Hz, + 0.2 dB; 
OFF: 0.05 Hz to 100 Hz, + 0.2 dB
• Gain settings
1, 2, 5, 10, 20 mm/mV + 25%
Wanded telemetry
• Frequency
Transmit: 69 kHz
Receive: 0–100 kHz
• Bandwidth
100 kHz
• Modulation
OOK
• Effective radiated power
–1.2 dB
µV/m
ZIP telemetry