Testing, Troubleshooting, and Maintenance
SENTINEL Meter Technical Reference Guide
7-11
Error Code
Error
Possible Cause(s)
Error Description
Err 1-----
Low Battery Error
Battery Voltage Low
Battery Connector Not 
Connected Properly
A low battery check is performed once a day and upon 
initialization. If a low battery level is detected during this 
procedure, the error code can be displayed in Normal Mode. 
This error can be programmed to continuously scroll during the 
display off time or lock on the meter display. A low battery 
continues to function; however, its reliability decreases over 
time.
If a dead battery is detected upon restoration of power, the error 
code will display until the battery is replaced. TOU and load 
profile functions will be disabled until the battery is replaced and 
the time is set. Total registers remain functional during a dead 
battery condition. However, TOU (rates A, B, C, D, E, F, G) 
registers and Load Profile Channels will not accumulate after 
an outage with a dead battery. If a dead battery level error 
occurs during normal operation, TOU and load profile will 
continue to operate until an outage occurs.
To retain data for TOU and Load Profile, the battery must be replaced with the power applied
to the meter.
Err -2----
Loss of Phase Error
The voltage on one of 
the phases dropped 
below 45 volts.
A Demand Reset after the voltage returns will clear this error.
Err --3---
Time-of-Use (TOU) 
Error
A problem was 
detected in the Time-
of-Use Schedule.
TOU functions have stopped because the meter could not 
interpret the schedule. To correct the problem, correct the 
problem in the TOU schedule and reprogram the meter.
Err ---4--
Reverse Direction 
Error
Reverse Power was 
detected on one or 
more phases.
A Reverse Direction error exists when the meter detects one 
second of (Active) energy on one or more phases in the reverse 
direction. This error is applicable to kWh only. Reverse direction 
detection is a selected feature during programming. This error 
is reported only when the meter is not measuring kWh received 
and/or kWh net.
The reverse direction error will clear when a demand reset is 
performed.
This error is not active with bidirectional measurement.
Err ----5-
Clock/Load Profile 
Error
Clock/Load Profile 
Error
This error occurs if the meter is powered up with a dead or no 
battery when it is configured for load profile. If this error occurs, 
the error code is displayed in Normal Mode. If a Clock/Load 
Profile error occurs, the load profile and TOU operations will be 
discontinued until the meter is reprogrammed. The error will 
then clear and the meter will resume normal operation.