49 E. Chase St.             Pensacola, FL 32502              tel 850.332-7900             fax 850-332-7904 
Device Serial Number Label 
Each ActiGraph device is equipped with a label stuck to it with pertinent device information inside the 
serial number.  For example the serial number MRA3A29120016 means: 
 
MRA3 = device model 
 
A = device model version 
 
29 = week of the year it was manufactured (29
th
 week)  
 
12 = year it was manufactured (2012) 
 
0016 = number of device manufactured that week 
Cleaning the Device 
ActiGraph devices are approved to be cleaned with a non-abrasive cloth and isopropyl alcohol.  The USB 
cap is to be inserted into the device and tightened prior to cleaning so no liquid can enter into the 
device and damage any of the internal electronics.  For optimal results, apply isopropyl alcohol to the 
cloth, clean the device for 15-20 seconds, and let air dry.  ActiGraph devices are not to be sterilized 
because damage may occur.  Though ActiGraph devices are not required to be cleaned on any 
predetermined schedule, it is advised that the devices be cleaned after each use by a user.  
Device Storage and Life Expectancy 
There is no specific requirement for storing ActiGraph devices but a dry location with temperature 
limitations between 0°C and 50°C is suggested.  For optimal battery life expectancy, routinely charge the 
device according to Table 4. 
The only limiting factor of an ActiGraph device when calculating life expectancy is the battery.  The 
battery has a life expectancy of five (5) years but can be replaced by ActiGraph personnel thus extending 
the life of the device.  The performance of the components inside the device does not degrade over 
time so the life expectancy of a device is immeasurable.   
Device Maintenance 
ActiGraph’s devices do not require routine maintenance or calibration.  Calibration is done during the 
manufacturing of the device and will last for the life of the device.  If maintenance is required on a 
device, or a piece of a device has gone missing, call ActiGraph Customer Support or visit the support 
section on the ActiGraph website 
 and initiate an RMA for your device.    
Any critical components that require maintenance are to be performed by trained ActiGraph personnel 
only.  All necessary training manuals, schematics, parts listings, and other necessary documentation, will 
be made available to the trained ActiGraph service technicians.   
Accessories 
Accessories sold at ActiGraph are constantly changing due to the ever changing features available with 
the wGT3X-BT device.  For a full list of current accessories available for the wGT3X-BT device, please visit 
ActiGraph’s website 
 .  Use the accessories