Mindray DS USA Inc. WT0707-00000 Manuel D’Utilisation
Physiological Alar
ms and E
vents
Alar
ms and Ev
en
ts
13 - 6
0070-10-0690-01
Panorama™ Operati
ng Ins
tructions
T Blood
Units: F°
Units: F°
2 or 3
Factory Default: 3
Factory Default: 3
Low Limit
N/A
Off, 80-100°
Factory Default: Off
Factory Default: Off
Off, 80-100°
Factory Default: Off
Factory Default: Off
Spectrum
™
/Spectrum OR
™
High Limit
N/A
Off, 95-110°
Factory Default: Off
Factory Default: Off
Off, 95-110°
Factory Default: Off
Factory Default: Off
T Blood
Units: C°
Units: C°
2 or 3
Factory Default: 3
Factory Default: 3
Low Limit
N/A
Off, 26-38°
Factory Default: Off
Factory Default: Off
Off, 26-38°
Factory Default: Off
Factory Default: Off
Spectrum/Spectrum OR
High Limit
N/A
Off, 35-43°
Factory Default: Off
Factory Default: Off
Off, 35-43°
Factory Default: Off
Factory Default: Off
NOTE:
Only available if Spectrum/Spectrum OR External Parameter Module (EPM) is installed and sensor is
attached
attached
ST Single
Units: mm
Units: mm
1, 2 or 3
Factory Default: 2
Factory Default: 2
High Limit
N/A
Off,
0.50 to10.0
Factory Default: Off
0.50 to10.0
Factory Default: Off
Off,
0.50 to 10.0
Factory Default: Off
0.50 to 10.0
Factory Default: Off
Passport 2
Spectrum/Spectrum OR
Telepack
Telepack
Low Limit
N/A
Off,
(-0.5) to
(-10.0)
Factory Default: Off
(-0.5) to
(-10.0)
Factory Default: Off
Off,
(-0.5) to
(-10.0)
Factory Default: Off
(-0.5) to
(-10.0)
Factory Default: Off
ST Dual
Units: mm
Units: mm
1, 2 or 3
Factory Default: 2
Factory Default: 2
High Limit
N/A
Off,
0.50-10.0
Factory Default: Off
0.50-10.0
Factory Default: Off
Off,
0.50-10.0
Factory Default: Off
0.50-10.0
Factory Default: Off
Spectrum/Spectrum OR
Telepack
Telepack
Low Limit
N/A
Off,
(-0.5) - (-10.0)
Factory Default: Off
(-0.5) - (-10.0)
Factory Default: Off
Off,
(-0.5) - (-10.0)
Factory Default: Off
(-0.5) - (-10.0)
Factory Default: Off
PVC
Units:
PVCs/min
Units:
PVCs/min
2 or 3
Factory Default: 3
Factory Default: 3
High Limit
N/A
Off, 1-30
Factory Default: Off
Factory Default: Off
Off, 1-30
Factory Default: Off
Factory Default: Off
Passport 2
Spectrum/Spectrum OR
Telepack
Telepack
Low Limit
N/A
N/A
N/A
Apnea Delay
Units: Seconds
Units: Seconds
2 or 3
Factory Default: 3
Factory Default: 3
High Limit
10-20
Factory Default: 15
Factory Default: 15
10-20
Factory Default: 15
Factory Default: 15
10-60
Factory Default: 60
Factory Default: 60
Passport 2
Spectrum/Spectrum OR
Low Limit
ALARM
ALARM
PRIORITY
ALARM LIMIT
NEONATE
PEDIATRIC
ADULT
DEVICES
N/A - Not Applicable