Mindray DS USA Inc. WT0707-00000 Manuel D’Utilisation

Page de 484
Panorama™ Operating Instructions
0070-10-0690-01
13 - 7
Alarms and Events
Physiological Alarms and Events
13.1.2
Physiological Alarm Event Messages
This section lists the alarm messages that may appear at the Panorama Central Station when 
an Physiological alarm event occurs.
PARAMETER
PANORAMA CENTRAL 
STATION MESSAGES
REASON 
Heart Rate (HR)
High Heart Rate Alarm
Low Heart Rate Alarm
Current value is either equal to or higher/lower 
than the set alarm limits.
Respiration (ECG 
Thoracic 
Impedance, CO
2,
 
and Gas Module)
High Respiration Rate 
Alarm
Low Respiration Rate 
Alarm
Current value is either equal to or higher/lower 
than the set alarm limits.
SpO
2
 (Masimo 
SET
®
, Nellcor
®
,
 
Nonin)
High SpO
2
 Alarm
Low SpO
2
 Alarm
Current value is either equal to or higher/lower 
than the set alarm limits.
IBPx (Sys) Labels   
include: IBP1, 
IBP2, IBP3, IBP4, 
ART, UA, LV, PA, 
CVP, ICP, LA, RA
High IBPx Systolic Alarm
Low IBPx Systolic Alarm
Current value is either equal to or higher/lower 
than the set alarm limits.
IBPx (Dia) Labels 
include: IBP1, 
IBP2, IBP3, IBP4, 
ART, UA, LV, PA, 
CVP, ICP, LA, RA
High IBPx Diastolic Alarm
Low IBPx Diastolic Alarm
Current value is either equal to or higher/lower 
than the set alarm limits.
IBPx (Mean) 
Labels include: 
IBP1, IBP2, IBP3, 
IBP4, ART, UA, 
LV, PA, CVP, ICP, 
LA, RA
High IBPx Mean Alarm
Low IBPx Mean Alarm
Current value is either equal to or higher/lower 
than the set alarm limits.
NOTE:
x = 1, 2, 3 or 4 (IBP3 and 4 are only available if the 
Spectrum/Spectrum OR External Parameter Module 
(EPM) is installed.)
INSP CO
2
 
(mmHg, %, kPa)
High CO
2
 Inspired Alarm
Current value is either equal to or higher than 
the set alarm limits.
ET CO
2
 
(mmHg,%, kPa)
High CO
2
 ET Alarm
Low CO
2
 ET Alarm
Current value is either equal to or higher/lower 
than the set alarm limits.
INSP O
2
O
2
 Inspired Alarm
There has been a violation of the O
2
 Inspired 
alarm.
ET O
2
O
2
 ET Alarm
There has been a violation of the O
2
 ET alarm.
INSP N
2
O
N
2
O Inspired Alarm
There has been a violation of the N
2
O Inspired 
alarm.
ET N
2
O
N
2
O ET Alarm
There has been a violation of the N
2
O ET alarm.
Agent
Agent Inspired Alarm
There has been a violation of the Agent Inspired 
alarm.
Agent ET Alarm
There has been a violation of the Agent ET 
alarm.
INSP Iso
Agent Iso Inspired Alarm
There has been a violation of the Iso Inspired 
alarm.
ET Iso
Agent Iso ET Alarm
There has been a violation of the Iso ET alarm.
INSP Enf
Agent Enf Inspired Alarm
There has been a violation of the Enf Inspired 
alarm.