Technical Alarms and Events
Alarms and Events
13 - 14
0070-10-0690-01
Panorama™ Operating Instructions
SpO
2
 No Sensor
The SpO
2
 Masimo sensor is not plugged in.
SpO
2
 Sensor Failed
Communication with the SpO
2
 Masimo sensor 
failed.
SpO
2
 Interference
There is interference in the communication with 
the SpO
2
 Masimo sensor.
SPO
2
 (NELLCOR) TECHNICAL EVENTS
SpO
2
 Motion
The SpO
2
 Nellcor sensor detects motion.
SpO
2
 Check Sensor
The SpO
2
 Nellcor sensor detects a problem.
SpO
2
 No Pulse
The SpO
2
 Nellcor sensor does not detect a pulse.
SPO
2
 (NONIN) TECHNICAL EVENTS
SpO2 (NONIN) Communication Error
There is interference in the communication 
between the SpO
source and the Telepack.
SpO
2
 Check Sensor
The SpO
2
 (Nonin) sensor detects a problem.
SpO
2
 No Sensor
The SpO
2
 (Nonin) sensor is not plugged in or the 
sensor is inoperable.
SpO
2
 Low Perfusion
The SpO
2
 (Nonin) sensor detects low perfusion.
SpO
2
 Pulse Search
The SpO
2
 (Nonin) sensor is attempting to detect a 
pulse.
IBPX TECHNICAL EVENTS
IBPx (Labels include: Art, UA, LV, PA, CVP, 
ICP, LA,RA) Not Calibrated/Zeroed
The IBP device is not calibrated or zeroed. 
NOTE:
x = 1, 2, 3 or 4 (IBP3 and 4 are only available if the 
Spectrum/Spectrum OR External Parameter Module 
(EPM) is installed)
CO
2
 TECHNICAL EVENTS
CO
2
 Not Calibrated/Zeroed
The CO
2
 device is not calibrated. 
CO
2
 Occlusion
The CO
2
 device detects a blockage in the filter 
line. 
O
2
 TECHNICAL EVENTS
O
2
 Not Calibrated/Zeroed
The O
2
 device is not calibrated. 
O
2
 Zeroing
The O
2
 device is zeroing.
O
2
 Occlusion
The O
2
 device detects a blockage in the filter 
line. 
N
2
O TECHNICAL EVENTS
N
2
O Not Calibrated/Zeroed
The N
2
O device is not calibrated. 
N
2
O Zeroing
The N
2
O device is zeroing. 
N
2
O Occlusion
The N
2
O device detects a blockage in the filter 
line. 
Technical Events (Continued)
MESSAGE
REASON
1
Event cannot be deleted from Event database
2
Event will not elicit technical event sound
3
Event will not display in Patient Status line
4
Event will always elicit the technical event sound regardless of the Technical Event sound setting
5
Event will post a Page request when the central station has the paging option enabled and the alarm is 
configured to send a page