Mindray DS USA Inc. WT0707-00000 Manuel D’Utilisation

Page de 484
Panorama™ Operating Instructions
0070-10-0690-01
13 - 15
Alarms and Events
Technical Alarms and Events
AGENT TECHNICAL EVENTS
Agent Not Calibrated/Zeroed 
The Agent device is not calibrated.
NIBP TECHNICAL EVENTS
NIBP Cuff Overpressure
The patient’s NIBP cuff has an overpressure 
episode. 
NIBP Unable to Measure
The patient’s NIBP cuff is unable to take a valid 
measurement. 
NIBP Retry
There is an NIBP measurement retry. 
NIBP Pump Higher
The patient’s NIBP cuff needs additional pressure 
to take a measurement. 
NIBP Measuring
The patient’s NIBP cuff is inflating.
NIBP Deflating 
The patient’s NIBP cuff is deflating.
CARDIAC OUTPUT TECHNICAL EVENTS
New CO Measurement 
2,3
The Cardiac Output run was accepted at the 
bedside monitor.
NOTE:
The CO parameter is only available when the Spectrum/
Spectrum OR External Parameter Module (EPM) is 
installed.
PAWP TECHNICAL EVENTS
New PAWP Measurement 
2,3
The PAWP value was accepted on the bedside 
monitor.
NURSE CALL TECHNICAL EVENTS
Nurse Call 
4
The Nurse Call button on the Telepack was 
selected. 
ATTENDANT PRESENT TECHNICAL EVENTS
Attendant Present 
2,3
The Attendant Present buttons on the Telepack 
were selected. 
BEDSIDE PRINTER TECHNICAL EVENTS
Printer out of paper
The bedside monitor local printer is out of paper. 
Printer Door is Open
The bedside monitor local printer has an open 
door condition. 
CENTRAL COMMUNICATION TECHNICAL EVENTS
Central Communications Lost
The connection between the ViewStation and the 
host central station was interrupted. 
Technical Events (Continued)
MESSAGE
REASON
1
Event cannot be deleted from Event database
2
Event will not elicit technical event sound
3
Event will not display in Patient Status line
4
Event will always elicit the technical event sound regardless of the Technical Event sound setting
5
Event will post a Page request when the central station has the paging option enabled and the alarm is 
configured to send a page