Innokas Yhtyma Oy VC150 Manuel D’Utilisation
6-10
VC150 Vital Signs Monitor
KO00065K
NIBP: NIBP modes of operation
Auto cycle determinations
Auto cycle mode automatically starts determinations at user-defined intervals.
In the auto cycle mode, the pressure must be below 5 mmHg (neonate) or 15
mmHg (adult) for at least 30 seconds before the next auto determination will be
started.
In the auto cycle mode, the pressure must be below 5 mmHg (neonate) or 15
mmHg (adult) for at least 30 seconds before the next auto determination will be
started.
1. Select the auto mode icon on home screen.
2. Select Cycle in the drop-down list.
3. Select the cycle length (measurement at every 2 min, 3 min, 4 min, 5 min, 10
min, 15 min, 20 min, 30 min, 60 min, 90 min and 120 min).
NOTE
Single determinations can be taken while in auto or profile cycle mode
without affecting the starting time for the next auto or profile determination.
However, venous return time will be honored before the next auto or profile
determination starts. If a single measurement lasts longer than when the
next auto cycle is supposed to start, the next auto cycle measurement is
skipped. You can also change the time interval while in auto cycle mode.
without affecting the starting time for the next auto or profile determination.
However, venous return time will be honored before the next auto or profile
determination starts. If a single measurement lasts longer than when the
next auto cycle is supposed to start, the next auto cycle measurement is
skipped. You can also change the time interval while in auto cycle mode.
4. Select Confirm to confirm the selection or select Cancel to cancel selection.
5. Select the green Cycle auto mode icon on the left side of the screen to start
the cycle. When a determination is complete, the time till the next
determination appears above the auto mode icon. If it becomes necessary
to stop the determination at any point, select the red Cycle auto mode icon.
determination appears above the auto mode icon. If it becomes necessary
to stop the determination at any point, select the red Cycle auto mode icon.