6-10
VC150 Vital Signs Monitor
KO00065K
NIBP: NIBP modes of operation
Auto cycle determinations
Auto cycle mode automatically starts determinations at user-defined intervals. 
In the auto cycle mode, the pressure must be below 5 mmHg (neonate) or 15 
mmHg (adult) for at least 30 seconds before the next auto determination will be 
started.
1. Select the auto mode icon on home screen.
2. Select Cycle in the drop-down list.
 
3. Select the cycle length (measurement at every 2 min, 3 min, 4 min, 5 min, 10 
min, 15 min, 20 min, 30 min, 60 min, 90 min and 120 min).
NOTE
Single determinations can be taken while in auto or profile cycle mode 
without affecting the starting time for the next auto or profile determination. 
However, venous return time will be honored before the next auto or profile 
determination starts. If a single measurement lasts longer than when the 
next auto cycle is supposed to start, the next auto cycle measurement is 
skipped. You can also change the time interval while in auto cycle mode.
4. Select Confirm to confirm the selection or select Cancel to cancel selection.
5. Select the green Cycle auto mode icon on the left side of the screen to start 
the cycle. When a determination is complete, the time till the next 
determination appears above the auto mode icon. If it becomes necessary 