Eno Scientific LLC WS310 Manuel D’Utilisation
MENU DESCRIPTIONS
PRODUCT AND VERSION – displays the product type, configuration and the software
version number.
version number.
SERIAL NUMBER – Displays the unit serial number.
SYSTEM TIME/DATE – Displays the system time and date that is used to tag logged
data. Press ENTER to set the year. Press ENTER again to set the month, etc. Press
ENTER after all have been set to return to menu mode.
data. Press ENTER to set the year. Press ENTER again to set the month, etc. Press
ENTER after all have been set to return to menu mode.
LANGUAGE – Changes the display language to english, spanish, or french.
UNITS – Selects the measurement system for the data. Press the UP or DOWN button
to scroll through the options: English for decimal feet, degrees Fahrenheit and gallons;
English ft-in for distance displayed as feet and inches; Metric for meters, degrees
Celsius, and liters, or Imperial for feet, degrees Fahrenheit and imperial gallons.
to scroll through the options: English for decimal feet, degrees Fahrenheit and gallons;
English ft-in for distance displayed as feet and inches; Metric for meters, degrees
Celsius, and liters, or Imperial for feet, degrees Fahrenheit and imperial gallons.
CONTRAST - Sets the contrast level on the LCD screen from 0 (brightest) to 10
(dimmest).
(dimmest).
COMMUNICATIONS - Changes the communication ports of the display unit.
RS232 SETUP – Enables continuous transmission of data on the RS232 port.
Press ENTER to choose set mode then UP/DOWN to select new setting.
Press ENTER to choose set mode then UP/DOWN to select new setting.
RS232 SETUP RATE – Changes the time between data updates on the RS232 port.
Press ENTER to choose set mode then UP/DOWN to select new setting from 0-60
minutes. If 0 is selected then every update is transmitted.
Press ENTER to choose set mode then UP/DOWN to select new setting from 0-60
minutes. If 0 is selected then every update is transmitted.
RS232 SETUP PORT – Changes the port configuration of the RS232 port.
Press ENTER to choose set mode then UP/DOWN to select new setting. Press
ENTER again to select the next parameter, etc. The default is 19,200 baud, 8 data bits,
1 stop bit, and no parity.
Press ENTER to choose set mode then UP/DOWN to select new setting. Press
ENTER again to select the next parameter, etc. The default is 19,200 baud, 8 data bits,
1 stop bit, and no parity.
RS485 SETUP PORT – Changes the port configuration of the RS485 port.
Press ENTER to choose set mode then UP/DOWN to select new setting. Press
ENTER again to select the next parameter, etc.
Press ENTER to choose set mode then UP/DOWN to select new setting. Press
ENTER again to select the next parameter, etc.
RF SETUP – Changes the RF channel used by the internal transceiver (11-26).
Press ENTER to choose set mode then UP/DOWN to select new setting.
Press ENTER to choose set mode then UP/DOWN to select new setting.
Eno Scientific Well Watch 310
Page 11