FACTORY RESET – Press ENTER to reset all settings back to factory defaults.
WELL WATCH SETUP – Selects settings to communicate with a remote Well Watch.
CONNECTION – Selects the type of connection the display unit will use to 
communicate with the remote Well Watch. Press ENTER to choose set mode then 
UP/DOWN to select new setting. Select RS485 if a wired connection is to be used. 
Wireless is selected as the default.
ADDRESS – Selects the address selected in the remote Well Watch. Press ENTER to 
choose set mode then UP/DOWN to select new setting (1-254). The default address is 
6 in both the display unit and the Well Watch.
RANGE MIN – Sets the minimum measurement range in the remote Well Watch. This 
requires that the connection with the remote Well Watch is already established. Press 
ENTER to choose set mode then UP/DOWN to select new setting.
RANGE MAX – Sets the maximum measurement range in the remote Well Watch. This 
requires that the connection with the remote Well Watch is already established.  Press 
ENTER to choose set mode then UP/DOWN to select new setting.
WELL TEMPERATURE – Sets the well temperature in the remote Well Watch. This 
requires that the connection with the remote Well Watch is already established.  Press 
ENTER to choose set mode then UP/DOWN to select new setting.
WELL DIAMETER – Sets the well diamater in the remote Well Watch. This requires 
that the connection with the remote Well Watch is already established.  Press ENTER 
to choose set mode then UP/DOWN to select new setting.
WELL DEPTH – Sets the well depth in the remote Well Watch. This requires that the 
connection with the remote Well Watch is already established.  Press ENTER to 
choose set mode then UP/DOWN to select new setting.
ALARM SETUP – Configures the alarm features of the display unit. An enabled alarm 
condition results in an audible alarm sound as well as a visual red indicator on the 
keypad. The audible alarm can be silenced with the ALARM key.
ALARM ENABLE – Selects the alarm type, High (water level is too close to the 
surface), Low (water level is too deep), or both. Press ENTER to choose set mode then 
UP/DOWN to select new setting.