Eno Scientific LLC WS310 Manuel D’Utilisation

Page de 21
The Well Watch 310 display unit contains an expanded memory which provides 
enough memory to record one data point per minute for over 50 years. This memory 
supports a windows compatible file system which can be accessed using the USB port. 
When the display unit is connected to a computer with the included USB cable, the 
computer will identify the unit as a mass storage device and show the available log files 
in an explorer window which can be copied directly to a directory on the computer and 
deleted from the memory. 
The log data files are simple ascii text files which can be opened with any common 
program such as Notepad, MS Word or Excel. The data files created are named 
WLOGxxx.TXT where xxx is the remote device address in the form 001 etc. Each data 
point stored is appended to the file corresponding to the currently installed remote 
device address. If a file does not exist for that address, it is created. Once the data is 
copied to the computer, the file can be deleted from the display unit. A new file will be 
created if the same address is used.
RS232 COMMUNICATIONS
The Well Watch 310 display unit offers a standard serial communications port for 
connection to a computer or remote device such as a SCADA system or remote 
readout. Screw terminals are provided for connection to the RS232 port. To connect to 
a standard device such as a PC with a standard 9 pin D connector, only 3 wires are 
necessary. The RX terminal on the well watch goes to pin 2 on the D connector, TX 
goes to pin 3 and GND goes to pin 5.  A simple serial to USB adapter is available from 
Eno Scientific or most computer stores to use with laptops without serial ports. The 
default serial configuration on the Well Watch is 19200 baud, 8 data bits, 1 stop bit, no 
parity, and no flow control. These settings can be changed from the menu as described 
earlier.
The serial port is bi-directional and allows communication from the display unit to the 
remote device and back. This allows some of the main settings to be changed from a 
remote location.  A standard terminal emulator program such as HyperTerminal 
(included with windows) or Tera Term Pro  (available for free download from 
) can be used from a computer to operate the display unit or 
view data remotely (see the section on remote operation). 
The Well Watch display can also be instructed to automatically transmit new data over 
the serial port as it is generated. Enable this feature from the RS232 SETUP menu as 
described earlier. 
  Eno Scientific  Well Watch 310
              Page  16