Page 29 / 83 
5.3  Aligning the guide bar with the direction of flow 
After adjusting the insertion distance and placing the flow meter in the desired position, tighten the 
cable gland to maintain the position. The orientation of the HydrINS 2/ HydrINS 2 Mini flow meter 
must now be adjusted. The head of the HydrINS 2 / HydrINS 2 Mini flow meter has a red arrow to 
orient the flow meter in the main direction of flow. Flow in this direction will count as positive, whilst 
flow in the reverse direction will count as negative. 
 
Figure 17: Position of the positive direction arrow on the probe head 
The guide bar supplied with the flow meter allows the electrodes to be positioned in the direction of 
flow. Insert the bar in its lodging in the flow meter head, and then match the arrow with the positive 
direction using the guide bar by placing it in the direction of flow. 
 
Figure 18: Aligning the flow meter in the direction of flow 
Once the direction and alignment have been adjusted, tighten the cable gland more tightly  to hold 
the flow meter firmly in its final measurement position. 
Main 
direction 
of 
flow 
(positive) 
Correct arrow direction 
 
Flow direction 
Guide bar direction