Rigol MSO1074Z-S 4-channel oscilloscope, Digital Storage oscilloscope, MSO1074Z-S Manuel D’Utilisation

Codes de produits
MSO1074Z-S
Page de 172
RIGOL 
DS1000Z Programming Guide 
2-103
 
:TRIGger:TIMeout (Option) 
 
Command List: 
 
 
:TRIGger:TIMeout:SOURce 
Syntax 
:TRIGger:TIMeout:SOURce <source> 
:TRIGger:TIMeout:SOURce? 
Description 
Set or query the trigger source in timeout trigger.   
Parameter 
Name 
Type 
Range 
Default 
<source> 
Discrete 
{CHANnel1|CHANnel2|CHANnel3|CHANnel4}  CHANnel1 
 
Return 
Format 
The query returns CHAN1, CHAN2, CHAN3 or CHAN4.   
Example 
:TRIGger:TIMeout:SOURce CHANnel2 
 
 
:TRIGger:TIMeout:SLOPe 
Syntax 
:TRIGger:TIMeout:SLOPe <slope> 
:TRIGger:TIMeout:SLOPe? 
Description 
Set or query the edge type in timeout trigger. 
Parameter 
Name 
Type 
Range 
Default 
<slope> 
Discrete 
{POSitive|NEGative|RFALl} 
POSitive 
 
Explanation 
  POSitive: start timing when the rising edge of the input signal passes through the 
trigger level.
 
  NEGative: start timing when the falling edge of the input signal passes through the 
trigger level.
 
  RFALl: start timing when any edge of the input signal passes through the trigger 
level.
 
Return 
Format 
The query returns POS, NEG or RFAL.   
Example 
:TRIGger:TIMeout:SLOPe NEGative