Lecroy WSXS-SPIBUS TD WSXS-SPIBUS TD trigger and decoder extension Compatible with LeCroy WaveSurfer® oscilloscopes. WSXS-SPIBUS TD Manuel D’Utilisation

Codes de produits
WSXS-SPIBUS TD
Page de 204
Serial Data Debug Solutions 
70 
919586 RevA 
Creating a CANbus Trigger Condition 
The CANbus Trigger dialog, with detail on some of the setup conditions, is shown in the following topics. 
Note: Refer to Using The D Supported Protocol Toolsets (on page 14) to correctly access the Trigger Condition 
dialog specific to your desired protocol. 
 
CANbus Trigger Setup Detail 
The following topic provides specific control settings for a CANbus Trigger. 
 
The previously numbered CANbus trigger sections correspond with the following explanations. 
1.
 
S
OURCE 
S
ETUP
 
 
DATA - The DATA field's pop-up dialog is used to select the appropriate channel or EXT input for each. Set 
this field up with caution or your trigger may not function correctly. Use the Threshold field to adjust the 
vertical level for the trigger. Much like an Edge trigger, a user must specify the level used in order to 
process the incoming signals and determine whether the desired serial data pattern is meeting the set 
trigger condition. 
2.
 
CAN
 
S
ETUP
 
 
Bitrate - Use the Bitrate field to adjust the value and match the bus to which you are connected. This 
bitrate selection is dynamically linked to the decoding bitrate (they are always the same value). Use the 
arrows to move through standard bit rates (10, 25, 33.333, 50, 83.333, 100, 125, 250, 500, and 1000 kb/s) 
and make a selection. Or, touch the number twice (with a finger, or using a mouse) and open a pop-up 
keypad and enter the value directly. 
3.
 
T
RIGGER 
T
YPE
 
Trigger Type - Depending on your Trigger Type selection, certain Frame ID and Data Pattern Setup fields 
are enabled or disabled as follows: 
 
 
All - Triggers on all signals. No Frame ID and Data Pattern ID Setup fields are enabled.