Remote - Only Frame ID Setup fields are enabled. 
 
Data - Both Frame ID and Data Pattern ID Setup fields are enabled. 
 
Error - Triggers only when an error signal occurs. No Frame ID and Data Pattern ID Setup fields are 
enabled. 
4.
 
S
ETUP 
F
ORMAT
 
Select either Binary or Hexadecimal (Hex) setup mode. The mode selected propagates through the entire 
CANbus trigger setup. 
Try selecting Binary mode, and set up the Frame ID in binary format, then re-select HEX mode and set up 
the data in hexadecimal format. Toggling back and forth between the modes does not result in loss of 
information. 
5.
 
F
RAME 
ID
 
S
ETUP
 
Frame ID Setup is used to trigger on a specific Frame ID value with either 11 or 29 Bits. 
When CANbus trigger selections are either Remote or Data, use the Frame ID Setup fields as follows: 
 
ID Condition - The ID condition can be set to many different values. If the ID condition is set to Equal
then a data definition can also be set. Any other ID condition precludes setting up a Data condition. 
The ID condition can be set to EqualNot EqualLess thanLess than or Equal toGreather thanGreater 
than or Equal to
In RangeOut Range, or Don't Care. a range, out of a range, or don’t care. 
 
ID Bits - The trigger can be set to trigger on CAN messages with either 11-bits (Standard CAN) or 29-bits 
(Extended CAN). You can also set the trigger so that it triggers on a message that meets a condition for 
either the 11-bit or 29-bit ID. For instance, there might be an 11-bit ID value that is present in both an 11-
bit and a 29-bit ID, and by choosing ALL, you could trigger when that ID is present on either of those 
messages. 
 
Frame ID - Specify the desired frame ID for triggering here. 
 
To Frame ID - When using an in range or out of range ID Condition (previous), specify a To Frame ID value 
for triggering. 
6.
 
D
ATA 
P
ATTERN 
S
ETUP
 
Fields on this section of the dialog are only enabled when using the Data trigger type. 
 
Data Condition - The Data Condition can be set to many different values. The Data condition can be set to 
EqualNot EqualLess thanLess than or Equal toGreather thanGreater than or Equal toIn Range, or 
Out Range.  
 
DLC - The DLC (data length code) can be set to any integer value from 0 to 8. It should match the DLC of 
the CAN message you want to trigger on. If you set it to a value less than 0, it defaults to 0. If you set it to 
a value greater than 8, it defaults to 8. 
 
Byte Order - Choose from either Motorola (default) or Intel byte orders. 
 
Start Bit and # Data Bits - The CANbus trigger allows you to trigger on up to 64 contiguous data bits (8 
data bytes). This maximum 64-bit string can start at any location in the CAN message data field - it is not 
limited to the start of a full byte or a nibble. 
The Start Bit can be any value from 0 to 63. If you enter a value less than 0, it defaults to 0. If you enter a 
value more than 63, it defaults to 63. The Start Bit value is always in LSB format (i.e., the bit number as 
shown on the decoded waveform, with bit 0 being at the far left and bit 63 being at the far right of the 
data string). Remember that the 1st data byte is bits 0-7, the 2nd data byte is bits 8-15, etc. Also, make 
sure that your Start Bit value makes sense in relation to the DLC Value. For instance, a Start Bit value of 32 
with a DLC Value of 4 is not going to result in a successful trigger.