Error Messages 
Test Errors 
The following error messages or messages may appear on the GPT-9600 
screen when configuring or running tests.   
 
Error Messages 
Description 
I ERR 
For ACW, DCW tests. Shown when the current 
is set too high. 
SHORT 
Voltage is too low. Indicates that the DUT 
could be shorted. 
V ERR 
For ACW, DCW and IR tests. Indicates that the 
voltage is too high.  
R ERR 
For IR tests. Resistance=0Ω. Check to see 
whether the DUT or test lead is shorting.  
TIME ERR 
For ACW tests. TIME ERR is displayed when 
the H setting is ≥15~20mA if the TIME setting 
is OFF. 
OVER 25W 
For DCW tests. OVER 25W is displayed if the H 
setting multiplied by the voltage setting is 
greater than 25W.