Gw Instek GPT-9612 Insulation measuring device, GPT-9612 Manuel D’Utilisation

Codes de produits
GPT-9612
Page de 75
 
OPERATION 
29 
Test Lead Connection 
This section describes how to connect the GPT-9600 to a DUT for 
withstanding or insulation resistance tests.  
ACW, DCW, IR Connection 
 
Background 
ACW, DCW and IR tests use the HIGH 
VOLTAGE terminal and RETURN terminal 
with the GHT-114 test leads.  
ACW, DCW, IR 
Connection 
High Voltage terminal
DUT
GPT-9000
Return terminal
 
 
Steps 
1.  Turn the power off on the safety tester. 
2.  Connect the high voltage test lead(red) to the 
HIGH VOLTAGE terminal and screw firmly 
into place. 
3.  Connect the return test lead(white) into the 
RETURN terminal and screw the protector bar 
into place, as shown below. 
 
HIGH VOLTAGE 
Terminal
RETURN 
Terminal