30
 
ACW and DCW Grounding Mode Information 
 
Background 
For ACW and DCW tests, the GROUND MODE 
is always set to ON. In the ON setting, the GPT-
9600 grounds the return terminal to the ground. 
This mode is best for DUTs that are grounded 
to an earth ground by their chassis, fixtures or 
operation environment. This mode measures 
the potential of the HIGH VOLTAGE terminal 
with respect to earth ground. This means that 
any stray capacitance/resistance that leaks to 
earth ground will also be measured. 
Example 
GROUND MODE = ON, DUT grounded 
High Voltage terminal
Return terminal
DUT
GPT-9600
stray 
resistance, 
capacitance
 
 