Beha Amprobe GT-800 PRO KITVDE-tester 4472081 Manuel D’Utilisation
Codes de produits
4472081
GT-600/GT-800 Instruction Manual
31
Power, Load Current Test (P/I
L
)
The power/load current test measures the apparent power S, active power P
A
,
mains voltage U
LN
, load current I
L
and power factor PF of the appliance,
connected to mains test/mains socket.
Warning
NEVER carry out this test unless you have first carried out a thorough
visual inspection, followed by a test of the protective bonding resistance
(for class I appliances), and then a test of insulation resistance and
protective conductor current or touch leakage test. You must verify that
these tests have passed before engaging this test.
(for class I appliances), and then a test of insulation resistance and
protective conductor current or touch leakage test. You must verify that
these tests have passed before engaging this test.
Caution
Live test! The appliance will be energized at mains voltage. For this
purpose, switch on the appliance. Appliances driven by motors or
equipped with heating units may present a danger for the person
testing (comply with the appliance instruction manual!). Ensure that the
appliance is in a safe condition to run and secure it prior to testing.
equipped with heating units may present a danger for the person
testing (comply with the appliance instruction manual!). Ensure that the
appliance is in a safe condition to run and secure it prior to testing.
Remarks:
Connect the appliance according to the figure 21.
If the load current is higher than 18.0 A for 10 seconds, the test will be
interrupted automatically and “RANGE OVERLOAD” message will be
displayed.
displayed.
Figure 21: Power / Load current test connection