Calibrate Screen
The OTP configuration can be saved to a file by clicking the Export button in the Data Memory window
and entering a file name. The OTP configuration will be saved to a *.gg file.
The module calibration data is also held in the bq27421 data Memory.
If the Gauge Dashboard is not displaying any information, then the bq27421 may not be supported by the
GaugeStudio version that is being used. A GaugeStudio upgrade may be required.
8
Calibrate Screen
To ensure proper calibration, perform the following steps. These steps may or may not be required,
depending on the type of calibration being performed. Only one calibration item can be selected and
calibrated at a time.
8.1
To Calibrate the bq27421
1. Select the types of calibration to be performed (see
).
2. Enter the measured values for the types selected.
3. Press the button to calibrate.
8.2
Voltage Calibration
Voltage calibration usually is not required. If needed, follow these steps:
1. Measure the voltage across Pack+ and Pack–.
2. Type the voltage value in mV into Enter measured value.
3. Press the Calibrate button.
8.3
Board Offset Calibration
This performs the offset calibration for the current offset of the board.
It is expected that no current is flowing through the sense resistor while performing this calibration step.
1. Remove load and short PACK– to LOAD–.
2. Press the Calibrate button.
12
bq27421 EVM: Single-Cell Impedance Track Technology
SLUUA63 – April 2013
Copyright © 2013, Texas Instruments Incorporated