Linear Technology DC1010A-A - LTC2493 24-Bit 4/2-ch I²C Delta Sigma ADC (Req. DC590) DC1010A-A DC1010A-A Fiche De Données

Codes de produits
DC1010A-A
Page de 5
QUICK START GUIDE FOR DEMONSTRATION CIRCUIT 1010 
2/4 CHANNEL 24,16,16/PGA-BIT ADC WITH I2C 
 
 
4
 
 
EXPERIMENTS
INPUT NOISE 
Solder a short wire between the CH0 and CH1 tur-
rets. Connect the inputs to ground through a short 
wire  and  start  taking  data.  LTC2493  Noise  should 
be approximately 0.12ppm of 5V (600nV RMS.) The 
electrical noise of the LTC2489 is also 600nV RMS, 
however this is masked by the 76.3µV quantization 
level.  If  the  input  is  well  between  code  transitions, 
the noise level will read zero. If the input is exactly 
between two codes such that the two adjacent out-
put codes have equal probability, the noise level will 
be  approximately  7.9ppm.  The  input  noise  of  the 
LTC2487 is apparent at very high gain settings (128 
or 256.) Note that with a 5V reference and gain set 
to  256,  1LSB  is  equal  to  298nV  –  which  is  lower 
than  the  600nVRMS  electrical  noise  of  the  input 
stage. 
COMMON MODE REJECTION 
Tie  the  two  inputs  (still  connected  together  from 
previous  experiment)  to  ground  through  a  short 
wire  and  note  the  indicated  voltage.  Tie  the  inputs 
to REF+; the difference should be less than 0.5 
µV 
due  to  the  140  dB+  CMRR  of  the  LTC2493.  The 
LTC2489 will produce less than 1LSB difference. 
INPUT NORMAL MODE REJECTION 
The  LTC2493  and  LTC2487  SINC4  digital  filter  can 
be software selected to reject 50Hz, 60Hz by 110dB, 
or  both  50  and  60Hz  by  87dB.  The  LTC2489’s 
SINC4  filter  is  fixed  at  50/60Hz.  To  measure  input 
normal  mode  rejection,  connect  COM  to  a  2.5  volt 
source such as an LT1790-2.5 reference or a power 
supply.  Connect any other input (CH0 – CH3) to the 
same  supply  through  a  10k
Ω  resistor.  Apply  a  10 
Hz, 2V peak-to-peak sine wave to the input through 
a 1uF capacitor.  
Set the rejection frequency to 55Hz (LTC2493 only) 
and  start  taking  data.  The  input  noise  will  be  quite 
large, and the graph of output vs. time should show 
large variations. 
Next,  slowly  increase  the  frequency  to  55  Hz.  The 
noise should be almost undetectable in the graph.