CLEANING, MAINTENANCE, REPAIR
Cleaning 
operations
Wipe the compressor housing down with a dry or slightly damp 
cloth.
Prepare all other parts for hygienic re-use in accordance with the 
instructions for use of the nebuliser.
Figure 5
WARNING: Danger of electric shock or malfunction caused by 
liquids getting into the compressor
Liquids can conduct electric current. If liquid has got into the com-
pressor, the housing may be live and a short circuit may occur in the 
device.
Never spray any liquids onto the compressor or into the compres-
sor's ventilation slits.
Wipe the compressor down with a damp cloth, not a wet one.
Do not immerse the compressor in liquids such as water or disin-
fectant solution.
Always switch the compressor off and disconnect the power 
adapter from the socket before working on the device.
If liquid does get into the compressor, the INQUA
®
 Inhalator must 
not be used under any circumstances. Contact the manufacturer 
(see back of instructions for use) or your sales partner before you 
use the INQUA
®
Inhalator again.
CAUTION: Danger of infection from contaminated nebuliser/
mask
Read the chapter entitled "Hygienic re-use" in the instructions for 
use of the nebuliser you are using, and carry out the procedures 
for hygienic re-use described there.