Inqua Inhalator 1000 BR030000 Fiche De Données

Codes de produits
BR030000
Page de 134
45
en
Maintenance
With the exception of the filter, the compressor contains no parts 
that are serviceable by the user.
If used twice daily for inhalation for a total of twice 20 minutes a day, 
the expected life cycle of the compressor is two years. If the com-
pressor is still being used after two years, contact INQUA or a local 
sales partner to have the device tested.
Filter change
The air filter in the compressor must be replaced after every 
200 operating hours, but at least once a year or if it becomes visibly 
dirty, discoloured or damp.
To do this, proceed as follows:
Unscrew the filter holder (7) from the compressor using a suitable 
coin.
Pull the old air filter (11) out of the filter holder and fit the new filter 
in its place.
Replacing the nebuliser
Note the information regarding material resistance and replaceabil-
ity in the instructions for use of the nebuliser.
Figure 6
Filter holder
11
Air filter
CAUTION: Damage to the compressor or impaired treatment 
caused by unsuitable air filter
Using an air filter made by a different manufacturer may damage the 
compressor or impair the effect of the inhalation treatment.
Use only air filters that are recommended by INQUA for use with 
the INQUA
®
Inhalator.