Leica CM1950 Manuel D’Utilisation

Page de 74
37
Leica CM1950 Cryostat
8.
Working with the Instrument
8.3
Configuring the Parameters
The instrument must be switched on at least 5 hours before the
planned use.
Turn the lamp on or off.
This button activates or deactivates the manual defrost of the cooling
chamber, specimen head or freeze shelf. (For more detailed handling
instructions, refer to the chapter on "Working with the Instrument" on
p. 39 of this Operating Manual.)
Pressing the key button for approx. 5 sec. locks the entire keyboard
(the LEDs in the clock go out).
Pressing the key button briefly, then the "-" key in the specimen head
control panel field, switches off the specimen head.
Pressing the key button briefly, then the "+" key in the specimen head
control panel field, switches the specimen head back on.
Programming the temperature of the cryochamber
The temperature of the cryochamber is set and indicated on the control
panel marked with the cryostat symbol.
The actual temperature is the standard indication.
Briefly pressing the "+" or "-" button displays the target temperature.
Set the desired value via the "+" / "-" buttons. Pushing the "+" or "-" button
for more than 1 sec. increases or decreases the chamber temperature
continuously.
The actual value will be indicated 5 seconds after finishing the program-
ming.
Refer to page 52
 
for a table with guide values. The temperature
values given there are based on long-term experience, but are
intended solely as guide values, as any tissue may require particu-
lar adjustments.