SSI America BINARY SOLVENT DELIVERY MODULE 90-2581 REV B Manuel D’Utilisation

Page de 48
 
 3-5 
2.6.2 Self-Flush 
Solution 
Self-flush heads require 250-500 mL of flushing solution. See 
section 1.1.4 for self-flush solution recommendations. A pH indicator 
that will indicate the concentration of salts in the solution is 
recommended as a reminder to change the solution. This flush solution 
should be replaced with a fresh solution weekly to avoid frequent 
pump maintenance. 
 
2.6.3  Inlet Tubing and Filters 
The table below shows the inlet tubing and filter used in the Binary 
Solvent Delivery Module. All inlet lines are supplied in a 30" (76 cm) 
length and are made of a fluoropolymer material. 
2.6.4  Priming the Pump and the Flushing Lines 
Connect a syringe to the outlet tubing. Run the pump at a flowrate 
of 3 to 5 ml/min. Prime the pump by pulling mobile phase and any air 
bubbles through the system and into the syringe (a minimum of 20 
ml). 
 
To prime the flush lines for a self-flush head, simply place the inlet 
line in the flush solution and connect a syringe to the outlet line and 
apply suction until the line is filled with flush solution.  Place the 
outlet line in the flush solution.  Secure both flush lines in the flush 
solution container so they stay immersed during pump operation. 
2.6.6  Long Term Pressure Calibration Accuracy 
This note applies if your pump is equipped with an electronic 
pressure transducer. The transducer has been zeroed and calibrated at 
the factory. Over the life of the pump, some drift may occur. For 
example, it is typical for the zero to drift < 10 p.s.i. after about 1 year 
of operation (i.e., with no back pressure on the pump a reading of 1-9 
p.s.i. may be displayed). A similar drift may also occur at higher 
pressures, and are typically less than 1% (e.g. <50 p.s.i. at 6,000 p.s.i. 
back pressure). 
 
If pressure calibration and/or drift is a concern, consult the factory. 
The pump can be shipped back to SSI for recalibration. Alternatively, 
written calibration and zero-reset procedures are available. Consult the 
factory to receive these instructions.