2.7 
Preparation for Storage or Shipping 
2.7.1 Isopropanol 
Flush 
Disconnect the outlet tubing from the pump. Place the inlet filter in 
isopropanol. Use a syringe to draw a minimum of 50 ml through the 
pump. Pump a minimum of 5 ml of isopropanol to exit. Leave the inlet 
tubing connected to the pump. Place the inlet filter in a small plastic 
bag and attach it to the tubing with a rubber band. Plug the outlet port 
with the shipping plug or leave a length of outlet tubing on the pump 
or cover the outlet port with plastic film. 
2.7.2  Packaging for Shipping 
 
CAUTION:  
Reship in the original carton, if possible.  If the original carton is 
not available, wrap the pump in several layers of bubble wrap and cushion the 
bottom, top, and all four sides with 2" of packaging foam.  Although heavy, an 
HPLC pump is a delicate instrument and must be carefully packaged to  
withstand the shocks and vibration of shipment. 
 