 1-4 
1.1.4  Self-Flush and Seal Life 
It is recommended that the Self Flush feature be used to improve 
seal life in a number of applications.  In particular, (as stated above) if 
pumping Buffers, Acids/Bases or any inorganic solution near 
saturation, the pump should utilize the Self Flush feature.  With every 
piston stroke, an extremely thin film of solution is pulled back past the 
seal.  If this zone is dry (without use of Self Flush), then crystals will 
form with continuous operation, which will ultimately damage the 
seal.   
 
Another application where Self Flush is highly recommended is 
when pumping Tetrahydrofuran (a.k.a. THR, Diethylene Oxide) or 
other volatile solvents such as acetone (Note: THF and most solvents 
are compatible only with all-Stainless Steel systems.  THF will attack 
PEEK).  Volatile solvents will dry rapidly behind the seal (without the 
use of Self Flush), which will dry and degrade the seal.   
 
IPA, Methanol, 20% IPA/water mix or 20% Methanol/water mix 
are good choices for the flush solution.  Consult the factory for 
specific recommendations.