2. INSTALLATION 
2.1 
Unpacking and Inspection 
Prior to opening the shipping container, inspect it for damage or 
evidence of mishandling. If it has been damaged or mishandled, notify 
the carrier before opening the container. Once the container is opened, 
inspect the contents for damage. Any damage should be reported to the 
carrier immediately. Save the shipping container. Check the contents 
against the packing list. 
2.2 Location/Environment 
The preferred environment for the Binary Solvent Delivery 
Module is normal laboratory conditions. The area should be clean and 
have a stable temperature and humidity. The specific temperature and 
humidity conditions are 10 to 30 °C and 20% to 90% relative 
humidity. The instrument should be located on a stable flat surface 
with surrounding space for ventilation and the necessary electrical and 
fluid connections. 
2.3 
Fluid Connections & Priming 
      There are only four connections to be made.   
1.  Connect left drawer pump solvent line supplied 
2.  Connect right drawer pump solvent line supplied 
3.  Connect line to injection valve from outlet “T”  
4.   Next, install the self flush.  Connect the self flush inlet and  
      outlet (opaque) tubing as shown on the the next page.   
 
 
 
 
 
 
Details on the proper installation of tubing and priming of the 
pump and self flush is shown on the next page.