18. ADJUSTMENT 
18.1 Adjustment of Measured Values
 
 
  Adjustment 
Adjustment comprises four kinds shown below. (1), (3) and (4) have already been adjusted. However, it is 
recommended for maintaining the measuring and printing accuracy to adjust them once a year. 
 
Calibration Details 
Reference
(1) Measured values adjustment 
Adjustment to keep the measured value against input within 
the accuracy ratings. 
Sec.18.1
(2) Shift programming of 
measured value 
Programming to shift a measured value 
Sec.18.2
(3) Adjustment of printing position 
Adjustment to set the printing range to the zero and span lines 
on the chart 
Sec.18.3
(4) Adjustment for time-axis 
synchronization of pens   
Adjustment of the gaps between pens for the time-axis 
synchronization 
Sec.18.4
 
1  
Adjustment of measured values
 
It is recommended for maintaining measuring and 
printing accuracy to adjust them once a year. 
(1) Execute the adjustment for each channel. 
(2) Execute the adjustment under the reference 
condition. (See the right table.) 
 
 
Items Reference 
conditions
Ambient temperature 23 
±2°C 
Ambient humidity 
55 
±10%RH 
Power voltage 
100 
±1VAC  
Power supply 
frequency 
50 or 60 
±0.5Hz 
2  
Preparation
 
1) Preparation of tools
 
Input types 
Tools 
DC voltage  Thermocouple
Resistance 
thermometer
Remarks 
DC standard 
voltage/current generator 
○ 
○ 
 
Accuracy: Should be better than 
±0.05%. 
Reference junction 
compensator 
 
○ 
 
0
°C ±0.2°C or less 
Thermocouple for test 
 
○ 
 
Same type of thermocouple as input type 
Standard variable resistor 
 
 
○ 
Accuracy: Should be better than 
±0.05%. 
3-core copper wire 
 
 
○ 
Same resistance for each 3 cores 
 
2) Connection
 
Connections depend upon the input types. See the next page. 
 
3) Before starting adjustment
 
(1) After the connection, mount the terminal board cover and turn on the power supply. 
(2) Before starting adjustments, warm up your recorder for at least 30 minutes until it becomes stable. (It is 
recommended to warm it up for at least one hour if possible.) 
 
 
Check and adjustments of the measured values need meticulous work in addition to the standard tools and 
reference conditions. 
If you need checks or adjustments of the measured values, please consult your nearest CHINO’s agent.
Reference conditions 
     
Adjustments 
Remarks