11.OTHER PROGRAMMING 
11.13 Input Filter 
 
The input filter has a function to stabilize the measuring input. This function can be programmed for each channel.   
Press       and        simultaneously for 3 seconds or more to display the [Input filter] programming screen. 
 
1  
Input Filter
 
A CR filter is mounted in the measuring circuit. In addition, a  
software filter (called as “input filter”) for the “primary delay 
computation” is also installed to smooth slight variations of the 
measuring input. The value for the programming is 
corresponding to “Time constant: T”. 
 
2  
Programming Flow Chart
  
<Example> To the time constant 5 seconds in CH2 
 
Operation screen 
[Check] 
 
(       +       ) 
    
3 sec. or more 
          
 
[To be ready for programming] 
 
          
 
[Programming channel]   
 
           
[Programming time constant]   
   
           
 
 
 
 
Error message appears? 
 
 
 
Temporarily stored 
 
Program other channels? 
 
 
      + 
Stored 
 
 
 
 
 
 
Press          or         to change 
channels. The input filter time constant of 
other channels can be checked.   
 
 
 
0, 1 to 10 
If “0” is programmed, the input filter is 
disabled. 
 
 
 
Store the [temporarily stored] 
parameters into memory. 
(A programming change mark is 
printed.) 
 
 
0.63 
Input filter: 0 (none) 
Input filter: T 
(T = 1 to 10) 
Input filter 
Channel 
YES 
NO 
Cursor
Cursor
Time constant
Cursor 
YES 
NO 
Programming range 
of time constant
 
CLEAR 
ALARM 
DISP
SET
END
2
TAG
ENTRY
ENTRY
SET
END
SHIFT
%
 
0
CLEAR
8
ALARM
  Storing 
Reference 3
   
Other channels check 
Reference 1
Reference 2
Ref. 2
Ref. 1
Keys other than 
SHIFT
 
°F
5
DATA I