14. PRINTING FORMAT
14.1
 Programming Automatic Range-Shift Printing 
 
This programming display only appears for the 
printing format (option). This programming is to 
change the trace printing range automatically. 
 
1  
Automatic Range-shift
 
The automatic range-shift function changes 
the trace printing range up to 5 stages 
according to the measured values. 
(1) This programming can be applied to each 
channel. 
(2) 
The total printing range can be 
programmed optionally irrespective of the 
programmed range in the [Range/Printing 
range]. 
(3) For each range, refer to the minimum 
printing range in section 22.1 (Input 
Specifications). The printing may be 
dispersed if each range is programmed to 
be less than the minimum printing range. 
(4) The range shifting is executed when the 
measured value exceeds approx. 0.5 mm 
from the minimum (zero) or the maximum 
(span) range. 
 
 
2  
Programming Mode
 
1) Screen 1 (Printing format check)
 
 
             
 
 
 
Dots appear at the selected format. 
By pressing      , the dots disappear and the 
cursor appears instead. 
 
2) Screen 2 (Programming channel)
 
 
 
  
 
 
By selecting automatic range-shift (), the cursor 
appears at the channel programming position. 
 
 
3) Screen 3 (Zero programming at No.1 range)
 
 
           
 
 
(1)  The default programmed value is the minimum 
value in the [Range/Printing range]. This 
programming can be changed as required. 
(2)   When all programming are completed, return to 
the above display and execute [Storing]. 
 
4) Screen 4 (Span programming at No. 1 to 5 
ranges)
 
 
         
 
 
(1)  
When the programming a span, the range 
number advances. Program the span up to the 
required range number.   
(2) After pressing       up to range No.5, press 
          again to return to Display 3. 
Mark        is the programming point and No. (1) to (5) shows 
the order of programming. By pressing        when the 
programming ends, the screen moves to the next 
programming screen. As No. 5 is not used in this example, 
only press        .
 
Printing range 
for 1st range
-100
400
(1)
(2)
Printing range 
for 2nd range
(3)
(400) 
600
Printing range 
for 3rd range
(4)
(600) 
800
Printing range 
for 4th range
(5)
(800) 
1200
-
100
400
600  800 1200
1st range 
(R1) 
4th range 
(R4) 
2nd
 range 
 
(R2) 
3rd   
range 
 
(R3) 
-200
1370
Measuring value 
<Example> Automatic range-shift programming
<Range No. 23 (K thermocouple)>
 
 
Standard 
Automatic 
range-shift 
Zone 
Compressed/expanded
Programming value 
(Max. 5 digits by left-justify)
Range No. 
Span 
Total printing 
range 
ENTRY
ENTRY
 
SET 
END 
ENTRY
ENTRY
 
Automatic range selection 
Channel 
Cursor 
Programming value 
(Max. 5 digits by left-justify)
Range No. 
Zero