Fluke Video Game Keyboard 175 Manuel D’Utilisation

Page de 18
Specifications
11
Specifications
Accuracy is specified for 1 yr after calibration, at operating temperatures of 18 
°
C to 28 
°
C, with relative humidity at 0 % to 95 %. Accuracy
specifications take the form of:
±
 ( [ % of Reading ] + [ Counts ] )
Maximum voltage between any terminal and earth ground:
1000 V DC or AC RMS
Surge Protection:
8 kV peak per IEC 61010
W
 Fuse for mA inputs:
W
 Fuse for A input:
440 mA, 1000 V FAST Fuse
11 A, 1000 V FAST Fuse
Display:
Digital: 6000 counts, updates 4/sec
Bar Graph:  33 segments;
Updates 40/sec
Frequency: 10,000 counts
Capacitance: 1,000 counts
Altitude:
Operating: 2000 m; Storage: 12,000 m
Temperature:
Operating: 
10 °C to +50 °C;
Storage: 
30 °C to +60 °C
Temperature coefficient: 0.1 X (specified accuracy / °C
(< 18 °C or > 28 °C)
Electromagnetic
Compatibility
 (EN 61326-1:1997):
In an RF field of 3 V/M, accuracy = specified accuracy except in temperature: specified accuracy 
±
 5 °C (9 °F)
Relative Humidity:
Noncondensing < 0 °C
0 % to 95 % @ 10 °C to 30 °C
0 % to 75 % @ 30 °C to 40 °C;
0 % to 40 % @ 40 °C to 50 °C
Battery Life:
Alkaline: 300 hrs typical
Size (H x W x L):
4.3 cm x 9 cm x 18.5 cm
Weight:
420 g
Safety Compliances:
ANSI/ISA S82.02.01, CSA C22.2-1010.1, IEC 61010 to 1000 V Overvoltage Category III, 600 V Overvoltage
Category IV
Certifications:
CSA, TÜV (EN61010), UL, 
P
;
 (
N10140),VDE