Specifications
13
Accuracy  ± ( [ % of Reading ] + [ Counts ] )
Function
Range 
1
Resolution
Model 175
Model 177
Model 179
DC Amps
4
60.00 mA
400.0 mA
6.000 A
10.00 A
0.01 mA
0.1 mA
0.001 A
0.01 A
1.0 % + 3
1.0 % + 3
1.0 % + 3
Hz
(AC- or DC- coupled,
V or A 
2, 3
 input )
99.99 Hz
999.9 Hz
9.999 kHz
99.99 kHz
0.01 Hz
0.1 Hz
0.001 kHz
0.01 kHz
0.1 % + 1
0.1 % + 1
0.1 % + 1
Temperature
-40 
°
C to +400 
°
C
-40 
°
F to +752 
°
F
0.1 
°
C
0.1 
°
F
NA
NA
1 % + 10
5
1 % + 18
5
MIN MAX AVG
For DC functions, accuracy is the specified accuracy of the measurement function 
±
 12 counts for changes longer than
275 ms in duration.
For AC functions, accuracy is the specified accuracy of the measurement function 
±
 40 counts for changes longer than 1.2 s
in duration.
1.  All AC voltage and AC current ranges are specified from 5 % of range to 100 % of range.
2.  Frequency is specified from 2 Hz to 99.99 kHz in Volts and from 2 Hz to 30 kHz in Amps.
3.  Below 2 Hz, the display shows zero Hz.
4.  Amps input burden voltage (typical): 400 mA input 2 mV/A, 10 A input 37 mV/A.
5.  Does not include error of the thermocouple probe.