Operation 
Hi Flow Sampler 
 
Instruction 0055-9017 
20 
3.14.2  Manual Measurement Mode 
 
3.14.2.1 
Manual 2-Stage Mode 
 
In the Manual 2-Stage mode the user starts a measurement at a high 
flow rate, waits for a stable leak rate reading, and then manually saves 
the readings using the Save function. The unit then automatically lowers 
the flow rate, at which time the user makes a second leak rate 
measurement. After waiting for the leak rate reading to once again 
stabilize, the user then manually saves all final measured and calculated 
measurements in memory. This mode is used when more control of the 
leak rate measurement process is required. 
 
Select Manual 2-Stage operation as follows: 
 
1.  From the Expanded Main Screen, select Menu > Operating Mode
 
1 1 / 2 2 / 0 4   0 9 : 4 5 : 3 0  
C a l i b r a t i o n
S e t   F l o w   U n i t s
M e n u   M o d e
O p e r a t i n g   M o d e
A c c e s s   R e c o r d s
[ 0 0 0 1 ]   ( M )
S - -
 
 
2.  Highlight Manual 2-Stage, and then press the I/O key. 
 
0 2 / 2 1 / 0 3   0 9 : 4 5 : 3 0  
A u t o m a t i c   2 - S t a g e
M a n u a l         2 - S t a g e
M a n u a l         1 - S t a g e
[ 0 0 0 5 ]   ( M )
S - -
 
 
3.  Press ESC to return to the Main Screen. Observe that (M) should 
appear at the bottom of the screen