Bacharach Hi Flow Sampler Manuel Du Propriétaire

Page de 72
Hi Flow Sampler 
Operation 
Instruction 0055-9017 
21
3.14.2.2 
Manual 1-Stage Mode 
 
In the Manual 1-Stage mode the user starts a measurement at a high 
flow rate and then waits for a stable leak rate reading to occur. During 
this time the flow rate can be changed using the Speed function on the 
Main Screen. After a stable leak rate reading is obtained, the user can 
manually save all final measured and calculated values in memory using 
the Save function on the Expanded Main Screen. This mode is used 
when more control of the leak rate measurement process is required. 
 
Select Manual 1-Stage operation as follows: 
 
1.  From the Expanded Main Screen, select Menu > Operating Mode
 
1 1 / 2 2 / 0 4   0 9 : 4 5 : 3 0  
C a l i b r a t i o n
S e t   F l o w   U n i t s
M e n u   M o d e
O p e r a t i n g   M o d e
A c c e s s   R e c o r d s
[ 0 0 0 1 ]   ( M )
S - -
 
 
2.  Highlight Manual 1-Stage, and then press the I/O key. 
 
0 2 / 2 1 / 0 3   0 9 : 4 5 : 3 0  
A u t o m a t i c   2 - S t a g e
M a n u a l         2 - S t a g e
M a n u a l         1 - S t a g e
[ 0 0 0 5 ]   ( M )
S - -
 
 
3.  Press ESC to return to the Main Screen. Observe that (M) should 
appear at the bottom of the screen