Bacharach Hi Flow Sampler Manuel Du Propriétaire

Page de 72
Hi Flow Sampler 
Calibration 
Instruction 0055-9017 
43
4 Calibration 
 
It is recommended that the Hi Flow Sampler be calibrated by your 
nearest Bacharach Service Center every 30 days. Calibration, however, 
can be performed in the field if your facility has the necessary equipment 
and qualified personnel to perform the procedures described in the 
following sections. 
 
The following procedures describe how to: 
 
  View the dates on which calibrations were performed 
  Verify calibration 
  Calibrate the background and leak gas sensors 
  Display the gain values of the calibration amplifier circuits  
  Erase calibration  
 
4.1 
View Calibration Dates 
 
Do the following to view the dates on which calibrations were performed 
on the background-gas and leak-gas sensors at both low (2.5% CH
4
) and 
high (100% CH
4
) calibration gas levels: 
 
1.  From the Main Screen select Menu > Calibration > Calibration 
Dates.  
 
2.  The following Calibration Dates Screen is displayed where mm/dd/yy 
is the month, day, and year calibration was performed:  
 
0 2 / 2 1 / 0 3     0 9 : 4 5 : 3 0  
B a c k   L o w     m m / d d / y y
B a c k   H i g h   m m / d d / y y
L e a k   L o w     m m / d d / y y
L e a k   H i g h   m m / d d / y y
[ 0 0 0 5 ]   ( A )
S - -
 
 
3.  Press the ESC key twice to return to the Main Screen. 
 
 
4.2 
Gas Calibration Equipment Setup 
 
The following equipment is recommended to both check the calibration 
and perform a calibration procedure on the Hi Flow Sampler: 
 
  Calibration Kit, P/N 0055-8002 
  Calibration Gas: 
 
2.5% CH
4
 in air, P/N 0051-1121 
 
100% CH
4
, P/N00 55-0060 
 
Zero Air, P/N 0051-7131 (optional) 
 
Connect the components of the calibration kit, calibration gas cylinder, 
and Hi Flow Sampler together as shown in Figure 4-1. DO NOT, however, 
attach the calibration equipment to the instrument until instructed to do 
so in the calibration procedure.