Bacharach Hi Flow Sampler Manuel Du Propriétaire

Page de 72
Hi Flow Sampler 
Calibration 
Instruction 0055-9017 
45
5.  Disconnect the hose from the BACKGROUND inlet port and connect 
it to the GAS inlet port. 
 
6.  Again wait several minutes for the gas reading to stabilize. At this 
time the gas concentration shown on the display should match the 
concentration stamped on the calibration gas cylinder. 
 
1 1 / 2 2 / 0 4   1 0 : 4 5 : 3 0
B a c k ( % )           0 . 0 0
L e a k ( % )           2 . 5 0
[ 0 0 0 5 ]   ( M )
S - -
 
 
7.  Disconnect the hose from the GAS port, and then allow the pumps to 
run until both the Back and Leak readings fall to zero percent. 
 
8.  Press the ESC key three times to return to the Main Screen. 
 
Note:
  If gas is still present in either sensor chamber when ESC is pressed, 
the instrument will automatically begin to purge the sensors until all gas is 
removed from the instrument. 
 
 
4.4 
Gas Calibration Procedure 
 
This procedure calibrates both the leak-gas sensor and the background-
gas sensor at 2.5% and 100% CH
4
. After setting up the calibration 
equipment as described in Section 4.2, proceed to calibrate the sensors 
as follows: 
 
1.  Before turning ON the instrument, be sure that the calibration 
equipment is not connected to the instrument’s GAS or 
BACKGROUND inlet port, and that the instrument is in an area of 
clean air. 
 
2.  Turn ON the instrument and wait its warm-up period to complete 
before proceeding to Step 3. 
 
3.  From the Main Screen select Menu > Calibration > Calibrate 
Sensors to display the Sensor Calibration Screen. 
 
0 2 / 2 1 / 0 3   0 9 : 4 5 : 3 0
B a c k g n d   @   2 . 5 %   C H 4
B a c k g n d   @   1 0 0 %   C H 4
L e a k   @   2 . 5 %   C H 4
L e a k   @   1 0 0 %   C H 4
[ 0 0 0 5 ]   ( A )
S - -
 
 
4.  Using the  keys, highlight the sensor to be calibrated along with its 
gas level. (In the example above, the background sensor will be 
calibrated using 2.5% CH
4
.)