Maintenance 
Hi Flow Sampler 
 
Instruction 0055-9017 
56 
5.7 
Flow Rate Verification 
 
This procedure verifies the instrument’s ability to provide sufficient 
airflow to capture the entire gas leak, and to verify the accuracy of the 
instrument’s calculated flow rate by using the optional external flow 
meter (P/N 0055-0270). 
 
Important:
 To assure accurate flow readings, conduct the following 
procedure in clean air, free of any combustible gases. 
 
1.  Install the flow meter onto the Hi Flow Sampler’s gas intake as shown 
in Figure 5-7. 
 
2.  Place the Hi Flow Sampler into its Manual 1-Stage Mode per 
Section 3.14.2.2. Then start the blower by selecting Start from the 
Main Screen.  
 
3.  Make note of the Flow(cfm) value as displayed on the Hi Flow 
Sampler’s LCD, and the fpm flow rate as displayed on the external 
flow meter. 
 
4.  Both Hi Flow Sampler's flow rate and the externally measured flow 
rate should intersect within the two lines of the graph shown in 
Figure 5-8. If the intersect point falls outside the two graph lines, 
then return the unit to your nearest Bacharach Service Center for 
evaluation. 
 
Figure 5-7. Flow Meter Installation