Maintenance 
Instruction 0055-9017 
57
 
 
Figure 5-8. Flow Rate Validation Graph 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
5.  With a fully charged battery pack, the measured flow rate should be 
greater than 9 cfm. If not, perform the following (see Figures 5-9 
and 5-10): 
 
  Clean the gas sample inlet screen.  
  Ensure that the exhaust port is not obstructed. 
  Remove the unit’s cover and ensure that the high- and low-
sampling point hoses are properly connected between the flow-
measurement tube and their corresponding solenoid valves, and 
that these hoses are not pinched or kinked. 
  Ensure that no obstruction is present inside the flow-
measurement tube. 
  If none of the above solves the problem, return the unit to your 
nearest Bacharach Service Center for evaluation.