Emerson MMI-20019043 Manuel D’Utilisation

Page de 374
Interaction between Density Damping and Added Damping
In some circumstances, both Density Damping and Added Damping are applied to the reported
density value.
Density Damping
 controls the rate of change in the density process variable. Added Damping
controls the rate of change reported via the mA output. If mA Output Process Variable is set to
Density
, and both Density Damping and Added Damping are set to non-zero values, density
damping is applied first, and the added damping calculation is applied to the result of the
first calculation.
4.5.4
Configure Density Cutoff
Display
Not available
ProLink II
ProLink
 > Configuration > Density > Low Density Cutoff
ProLink III
Device Tools
 > Configuration > Process Measurement > Density
Field Communicator Configure > Manual Setup > Measurements > Density > Density Cutoff
Overview
Density Cutoff
 specifies the lowest density value that will be reported as measured. All
density values below this cutoff will be reported as 0.
Procedure
Set Density Cutoff to the value you want to use.
The default value for Density Cutoff is 0.2 g/cm
3
. The range is 0.0 g/cm
3
 to 0.5 g/cm
3
.
Effect of Density Cutoff on volume measurement
Density Cutoff
 affects liquid volume measurement. If the density value goes below Density
Cutoff
, the volume flow rate is reported as 0. Density Cutoff does not affect gas standard
volume measurement. Gas standard volume values are always calculated from the value
configured for Standard Gas Density.
4.6
Configure temperature measurement
The temperature measurement parameters control how temperature data from the
sensor is reported. Temperature data is used to compensate for the effect of temperature
on the sensor tubes during flow measurement .
The temperature measurement parameters include:
Temperature Measurement Unit
Temperature Damping
Configure process measurement
50
Micro Motion
®
 Model 2700 Transmitters with Analog Outputs