Page de 374
4.6.1
Configure Temperature Measurement Unit
Display
OFF-LINE MAINT
 > OFF-LINE CONFG > UNITS > TEMP
ProLink II
ProLink
 > Configuration > Temperature > Temp Units
ProLink III
Device Tools
 > Configuration > Process Measurement > Temperature
Field Communicator Configure > Manual Setup > Measurements > Temperature > Temperature Unit
Overview
Temperature Measurement Unit
 specifies the unit that will be used for temperature
measurement.
Procedure
Set Temperature Measurement Unit to the option you want to use.
The default setting is Degrees Celsius.
Options for Temperature Measurement Unit
The transmitter provides a standard set of units for Temperature Measurement Unit. Different
communications tools may use different labels for the units.
Options for Temperature Measurement Unit
Table 4-12:   
Unit description
Label
Display
ProLink II
ProLink III
Field Commu-
nicator
Degrees Celsius
°C
degC
°C
degC
Degrees Fahrenheit
°F
degF
°F
degF
Degrees Rankine
°R
degR
°R
degR
Kelvin
°K
degK
°K
Kelvin
4.6.2
Configure Temperature Damping
Display
Not available
ProLink II
ProLink
 > Configuration > Temperature > Temp Damping
ProLink III
Device Tools
 > Configuration > Temperature
Field Communicator Configure > Manual Setup > Measurements > Temperature > Temp Damping
Configure process measurement
Configuration and Use Manual 
51