5.5.1
Configure Fault Timeout
Display
Not available
ProLink II
ProLink
 > Configuration > Analog Output > Last Measured Value Timeout
ProLink
 > Configuration > Frequency/Discrete Output > Frequency > Last Measured Value Timeout
ProLink III
Device Tools
 > Configuration > Fault Processing
Field Communicator Configure > Alert Setup > Alert Severity > Fault Timeout
Overview
Fault Timeout
 controls the delay before fault actions are performed.
Restriction
Fault Timeout
 is applied only to the following alarms (listed by Status Alarm Code): A003, A004, A005,
A008, A016, A017, A033. For all other alarms, fault actions are performed as soon as the alarm is
detected.
Procedure
Set Fault Timeout as desired.
The default value is 0 seconds. The range is 0 to 60 seconds.
If you set Fault Timeout to 0, fault actions are performed as soon as the alarm condition is
detected.
The fault timeout period begins when the transmitter detects an alarm condition. During
the fault timeout period, the transmitter continues to report its last valid measurements.
If the fault timeout period expires while the alarm is still active, the fault actions are
performed. If the alarm condition clears before the fault timeout expires, no fault actions
are performed.
Tip
ProLink II allows you to set Fault Timeout in two locations. However, there is only one parameter, and
the same setting is applied to all outputs.
5.5.2
Configure Status Alarm Severity
Display
Not available
ProLink II
ProLink
 > Configuration > Alarm > Severity
ProLink III
Device Tools
 > Configuration > Alert Severity
Field Communicator Configure > Alert Setup > Alert Severity > Set Alert Severity
Configure device options and preferences
90
Micro Motion
®
 Model 2700 Transmitters with Analog Outputs