Error Log Messages
44
Chapter 4    Configuring the Monitor    
Error Log Messages
Table 4-5Error Log Messages
Message
Location of Error and Solution
Err 18
Incorrect type of paper loaded (same as 
Err PAP
). 
Replace with recommended paper.
Err 19
Battery low or empty of charge (same as 
Err bAt
). 
Change the batteries as soon as possible.
Err 70
Modem not responding. 
Modem is either not installed, incorrectly installed, faulty or not an approved 
model.
Err 77
Error during transmission of data. 
Probably due to external sources. Retransmit the data.
Err 500
General failure - no specific location. 
See the Troubleshooting Flowchart in Chapter 8.
Err 501, 
513
Front End Board (Single Ultrasound model). 
See the Troubleshooting Flowchart in Chapter 8.
Err 502
Power Supply Board. 
See the Troubleshooting Flowchart in Chapter 8.
Err 503
CPU Board. 
See the Troubleshooting Flowchart in Chapter 8.
Err 510
Recorder Board. 
See the Troubleshooting Flowchart in Chapter 8.
Err 511
Front End Board (Dual Ultrasound model). 
See the Troubleshooting Flowchart in Chapter 8.
Err 512
Intrapartum Board. 
See the Troubleshooting Flowchart in Chapter 8.
Err 531
Combined Interface Board. 
See the Troubleshooting Flowchart in Chapter 8.
Err 532
Modem Interface Board. 
See the Troubleshooting Flowchart in Chapter 8.
Err 601
 Paper 
feed. 
See the Troubleshooting Flowchart in Chapter 8.
Err 602
 
Incorrect type of paper. 
Load paper with the correct scale or change the setting.
 
Err 610
 
No loudspeaker, or fuse has blown. 
See the Troubleshooting Flowchart in Chapter 8.