Error Log Messages
Chapter 4 Configuring the Monitor
45
Err 611
Wrong loudspeaker impedance.
See the Troubleshooting Flowchart in Chapter 8.
See the Troubleshooting Flowchart in Chapter 8.
Err 800
Software error: checksum wrong or download incomplete.
Reload the software.
Reload the software.
Err 820
Upgrade Key is for a different type of monitor.
Use the correct key.
Use the correct key.
Err 888
No software upgrade done.
Disconnect transducers and telemetry first.
Disconnect transducers and telemetry first.
Table 4-5Error Log Messages
Message