GE Q HP Manuel D’Utilisation

Page de 20
p. 4   
 
Table 1. HiTrap SP HP and HiTrap Q HP columns characteristics
Column volumes 
1 ml or 5 ml
Column dimensions 
0.7 × 2.5 cm (1 ml) and 1.6 × 2.5 cm (5 ml)
Total ionic capacity 
0.14–0.20 mmol (Cl
)/ml medium (Q)
 
0.15–0.20 mmol (H
+
)/ml medium (SP)
Dynamic binding capacity 
SP: approx. 55 mg ribonuclease/ml medium
 
(0.1 M sodium acetate, pH 6.0 at 1 ml/min)
 
Q: approx. 50 mg HSA/ml medium (20 mM Tris-HCl,
 
pH 8.2 at 1 ml/min)
Mean particle size 
34 μm
Bead structure 
6% highly cross-linked spherical agarose
Maximum flow rates 
HiTrap 1 ml: 4 ml/min, HiTrap 5 ml: 20 ml/min
Recommended flow rates 
HiTrap 1 ml: 1 ml/min, HiTrap 5 ml: 5 ml/min
Maximum backpressure 
0.3 MPa, 3 bar, 42 psi
Chemical stability 
All commonly used buffers
Charged group 
SP: – CH
2
CH
2
CH
2
SO
3
 Q: 
– 
CH
2
N
+
(CH
3
)
3
pH stability*
Short term 
SP: 3–14, Q: 1–14
Working 
SP: 4–13, Q: 2–12
Long term 
SP: 4–13, Q: 2–12
Storage temperature 
+4° to +30 °C
Storage buffer 
SP: 20% ethanol, 0.2 M sodium acetate
 
Q: 20% ethanol
Avoid 
SP: Oxidizing agents, cationic detergents and  
 buffers
 
Q: Oxidizing agents, anionic detergents and   
 buffers
*  The ranges given are estimates based on our knowledge and experience.
Please note the following:
 
pH stability, long term refers to the pH interval where the medium is stable over 
a long period of time without adverse eff ects on its subsequent chromatographic 
performance.
 
pH stability, short term refers to the pH interval for regeneration, cleaning-in-place and 
sanitization procedures.