Pressure Systems 9046 Manuel D’Utilisation

Page de 72
Pressure Systems, Inc.                                                                                        Model 9046 User’s Manual 
 
 
 
Page 57 
 
www.PressureSystems.com 
5.7.6  Command ’c’ - Sub-command Index 05: Select Prefix/Data Groups in a 
Stream 
This sub-command sets additional options that cause an already defined stream to deliver a 
specific prefix and/or other groups of data to the host. By default, only the Primary Transducer 
(i.e., Temperature/ Resistance) E. U. data (and no prefix) are delivered for the channels and 
format
 currently specified by the pre-requisite "c 00" sub-command for that stream.
 
 
Command 
 
c 05 st bbbb” 
 
‘c’  is the operation code 
‘05’ is the sub-command index (‘ii’) for 
Select 
Prefix/Data Groups
st’ is the stream id digit (1, 2, or 3 only)  
‘ bbbb’ is a hex option field (bit map) to select which 
prefix/data group options will be returned in the data 
stream (see table below) 
NOTE: all parameters are separated by a 
space 
character
 
Response 
 
A” or “Nxx” 
‘A’ for acknowledge; ‘N’ for negative acknowledge, 
error code xx
 
Description
: If this sub-command is never executed for a particular stream, then Primary 
Transducer (i.e., Temperature/Resistance)
 EU Data are delivered (by default) in that stream 
following the fixed format binary header (tssss as described by the “01” sub-command). 
However, this sub-command may also delete these default data from a stream (by not 
specifying them) as well as add other selected acquired data groups to a stream (by specifying 
them).  
The bit map values (shown in the following table) may be added together to specify all the 
actual data groups that will be delivered in each scan of the specified stream.  The first two table 
entries, if their “bits” are specified,  will cause two-byte binary (16-bit, big endian) status values 
to be delivered in the stream scan (immediately following the binary stream header).  The third 
table entry, if specified, will cause the Primary Transducer (i.e., Temperature/Resistance) EU 
Data
 to be delivered (next), per the specified pre-requisite ‘c 00’ sub-command’s format (f), and 
for just the channels specified  in that sub-command. The remaining table entries  will cause 
other special data groups (i.e., Primary Transducer data in counts or volts,  and Other 
Transducer (UTR Temp./Volts) in EU, A/D counts, or voltage forms) to also be delivered in each 
stream scan.  Each of these special data groups is also output  in the order of its table entry 
(within the scan) if its “bit” is specified.  Each group will also have a datum per the specified 
channels, and be in the specified format.