Advanced Entry Calibration
Advanced Entry Calibration is the alternate method for special-case media
types. The procedure has two parts:
1) Using different sampling algorithms, displayed sensor readings have to be
listed for each label and TOF value.
2) From this data, the best algorithm is chosen; then those values are manually
entered for the ‘Paper’, ‘Gap’ (or ‘Mark’), and ‘Empty’ variables.
+
Advanced Entry Calibration will override all previous calibration
settings; this method should be used only when Standard Calibration
has failed.
settings; this method should be used only when Standard Calibration
has failed.
To perform an Advanced Entry Calibration:
Step
Operator Action
Displayed Message
Comment
1
(Media should be loaded
and the appropriate
Sensor Type selected.)
and the appropriate
Sensor Type selected.)
Turn ‘On’ the printer.
Wait until the
printer initializes
(about six
seconds).
(about six
seconds).
2
Press the
key.
Raise the printhead
assembly.
assembly.
You are entering
Menu Mode.
3
Press the key to
enter the Media Settings
menu.
enter the Media Settings
menu.
See Section 4.1.2
for menu layout.
4
Press the key
and scroll to ‘Sensor
Calibration’.
and scroll to ‘Sensor
Calibration’.
Press the
key to abort this
procedure.
5
Press the
key.
You are entering
the calibration
submenu.
submenu.